Specifications: Fluke Calibration 4180/4181 Precision IR Calibrators

Specifications 4180 4181
Temperature range (@ 23°C ambient, 0.95 emissivity) –15°C to 120°C 35°C to 500°C
Display accuracy1 ±0.40°C at –15°C
±0.40°C at 0°C
±0.50°C at 50°C
±0.50°C at 100°C
±0.55°C at 120°C
±0.35°C at 35°C
±0.50°C at 100°C
±0.70°C at 200°C
±1.20°C at 350°C
±1.60°C at 500°C
Stability ±0.10°C at –15°C
±0.05°C at 0°C
±0.10°C at 120°C
±0.05°C at 35°C
±0.20°C at 200°C
±0.40°C at 500°C
Uniformity (5.0 in dia of center of target)2 ±0.15°C at –15°C
±0.10°C at 0°C
±0.25°C at 120°C
±0.10°C at 35°C
±0.50°C at 250°C
±1.00°C at 500°C
Uniformity (2.0 in dia of center of target)2 ±0.10°C at -15°C
±0.10°C at 0°C
±0.20°C at 120°C
±0.10°C at 35°C
±0.25°C at 250°C
±0.50°C at 500°C
Heating time 15 min: –15°C to 120°C
14 min: 23°C to 120°C
45 min: 35⁰ C to 500⁰ C
Cooling time 15 min: 120°C to 23°C
20 min: 23°C to –15°C
100 min: 500°C to 35°C
40 min: 500°C to 100°C
Stabilization time 10 minutes 10 minutes
Nominal emissivity3 0.95 0.95
Thermometer emissivity compensation 0.9 to 1.0 0.9 to 1.0
Target diameter 152.4 mm (6 in) 152.4 mm (6 in)
Computer interface RS-232 RS-232
Power 115 V AC (±10%), 6.3 A,
50/60 Hz, 630 W
230 V AC (±10%), 3.15 A,
50/60 Hz, 630 W
115 V AC (±10%), 10 A,
50/60 Hz, 1000 W
230 V AC (±10%), 5 A,
50/60 Hz, 1000 W
Fuse(s) 115 V AC 6.3 A, 250 V, slow blow
230 V AC 3.15 A, 250 V, T
115 V AC 10 A, 250 V, fast blow
230 V AC 5 A, 250 V, F
Size (H x W x D) 356 x 241 x 216 mm
(14 x 9.5 x 8.5 in)
356 x 241 x 216 mm
(14 x 9.5 x 8.5 in)
Weight 9.1 kg (20 lb) 9.5 kg (21 lb)